Retinoscopy Simulator

Retinoscopy Simulator

Retinoscopy (Ret) is a technique to obtain an objective measurement of the refractive error of a patient’s eyes. The examiner uses a retinoscope to shine light into the patient’s eye and observes the reflection (reflex) off the patient’s retina. While moving the streak or spot of light across the pupil the examiner observes the relative movement of the reflex then uses a phoropter or manually places lenses over the eye (using a trial frame and trial lenses) to “neutralize” the reflex.

Retinoscopy Simulator
Retinoscopy Simulator

Static retinoscopy is a type of retinoscopy used in determining a patient’s refractive error. It relies on Foucault’s principle, which states that the examiner should simulate optical infinity to obtain the correct refractive power. Hence, a power corresponding to the working distance is subtracted from the gross retinoscopy value to give the patient’s refractive condition, the working distance lens being one which has a focal length of the examiner’s distance from the patient (e.g. +1.50 dioptre lens for a 67 cm working distance).Myopes display an “against” reflex, which means that the direction of movement of light observed from the retina is a different direction to that in which the light beam is swept. Hyperopes, on the other hand, display a “with” movement, which means that the direction of movement of light observed from the retina is the same as that in which the light beam is swept.

Static retinoscopy is performed when the patient has relaxed accommodative status. This can be obtained by the patient viewing a distance target or by the use of cycloplegic drugs (where, for example, a child’s lack of reliable fixation of the target can lead to fluctuations in accommodation and thus the results obtained). Dynamic retinoscopy is performed when the patient has active accommodation from viewing a near target.

Retinoscopy is particularly useful in prescribing corrective lenses for patients who are unable to undergo a subjective refraction that requires a judgement and response from the patient (such as children or those with severe intellectual disabilities or communication problems). In most tests however, it is used as a basis for further refinement by subjective refraction. It is also used to evaluate accommodative ability of the eye and detect latent hyperopia.

This shows you how to perform retinoscopy using the SPH/CYL method.
1. Neutralize the MORE plus meridian with a SPH lens (remember: fix all with motion).
2. Leave the SPH lens on the eye and rotate the beam 90 degrees away.
3. Neutralize this one with a minus CYL lens, the axis of this CYL should be aligned with the axis of your beam.
4. Recording: the SPH – your working distance, the cylinder power and axis as it is on the eye
5. Be happy Retinoscopy Simulator smile.gif

 

Source:en.wikipedia.org/wiki/Retinoscopy

eyeontechs.com/new/wp-content/uploads/2009/04/retinoscopysimulator.swf

Retinoscopy Simulator

Retinoscopy Simulator

Ophthalmology Notes

We do our best to simplify ophthalmology in Short Notes and exchange practical experiences between ophthalmologists all over the world.

Recently Published Stories

Albino Fundus with Nystagmus Foveal Hypoplasia(By MII Ret cam )

MII Ret Cam Pictures in Multiple Retinal Diseases ( Central and Peripheral)

MII Ret Cam : Make In India Retinal Camera is the First Smartphone Based Fundus imaging device with Capability of

Colorblindness Glasses

Colorblindness Glasses

Colorblindness Glasses A person with red-green color blindness (more accurately called color vision deficiency) experiences the world differently because their red

Pixium Vision Implants Restore Sight

Pixium Vision Implants Restore Sight

Pixium Vision Implants Restore Sight Prima Vision, a company based in Paris, France, is reporting that its PRIMA wireless subretinal

KAMRA Inlay Corneal Implant to Replace Reading Glasses

KAMRA Inlay Corneal Implant to Replace Reading Glasses

KAMRA Inlay Corneal Implant to Replace Reading Glasses Reading glasses are a necessity for most people older than 40, when

Multifocal IOL evolution

Multifocal IOL evolution

Multifocal IOL evolution By Cheryl Guttman Krader; Reviewed by Rudy M.M.A. Nuijts, MD, PhD, and Soraya M.R. Jonker, MD Maastricht,

LENS IMPLANT COULD IMPROVE VISION BEYOND 20/20

LENS IMPLANT COULD IMPROVE VISION BEYOND 20/20

LENS IMPLANT COULD IMPROVE VISION BEYOND 20/20 A new bionic lens implant could give patients the unique ability to see

"MII Ret Cam" First World Smartphone Based Peripheral Fundus imaging device

“MII Ret Cam” First World Smartphone Based Peripheral Fundus imaging device

“MII Ret Cam” – World First Smartphone Based Fundus imaging device with Capability of Imaging Peripheral Retina” Smartphone has revolutionized

First Open Source Smartphone Retinal Camera

Smartphone Retinal Camera

Smartphone Retinal Camera Inventor names: David Myung, Alexandre Jais, Robert Chang, and Mark Blumenkranz from Stanford University. Introduction Image-capture in

Kestenbaum’s rule (Strength of reading plus lens)

Kestenbaum’s rule (Strength of reading plus lens)

Kestenbaum’s rule (Strength of reading plus lens) Kestenbaum’s rule: used to estimate strength of plus lens required to read newspaper

iExaminer for Fundus Exams

iExaminer for Fundus Exams

iExaminer for Fundus Exams The iExaminer from Intuitive Medical Technologies in Shreveport, Louisiana is a simple iPhone 4 attachment for the popular